Transducerless time domain reflectance measurement of semiconductor thermal properties

Abstract

The dominant techniques for laser-based thermal measurement, time-domain thermoreflectance (TDTR) and frequency-domain thermoreflectance, generally require that samples be coated with a metal transducer layer. The properties of the transducer and its interface to the sample are then important to the interpretation of results, and heat conduction in the transducer layer can limit measurements of ultra-thin samples or in-plane conductivity. We demonstrate the feasibility of measuring semiconductor samples using standard TDTR instrumentation, without the use of a metal transducer. A mathematical model accounting for non-zero laser penetration depth as well as both thermal and charge carrier effects is presented, and the ability of the model to describe and fit to experimental data for silicon and germanium samples is demonstrated. Though at this stage measurement uncertainties are larger than for traditional TDTR, we anticipate that the use of this technique will expand the range of samples that can be measured with existing TDTR setups.

Document Details

Document Type
Pub Defense Publication
Publication Date
Jan 01, 2022
Source ID
10.1063/5.0069360

Entities

People

  • Junqiao Wu
  • Sarah Warkander

Organizations

  • Lawrence Berkeley National Laboratory
  • United States Department of Defense
  • United States Department of Energy

Tags

Readers

  • Acoustical Oceanography.
  • Computational Modeling and Simulation
  • Materials Science and Engineering.

Technology Areas

  • Directed Energy
  • Directed Energy - Pulsed-Laser Deposition
  • Microelectronics
  • Microelectronics - Graphene