STANDARD VALUES AND VARIATIONS FOR PAPER STRIP ELECTROPHORESIS
Abstract
Results of an 8 x 8 Latin square design controlling several variables in paper strip electrophoresis (Spinco procedure) indicates that determinations of a single trial or run are unique. Hence, care must be used in comparing determinations obtained from two different runs. Within a single run, there is no evidence that the position of any strip on the apparatus results in an obvious bias. An analysis of variance of these data yields estimates of the standard deviations associated with each serum protein component for several anticipated comparisons. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- May 02, 1961
- Accession Number
- AD0258361