RESEARCH INVESTIGATION OF P-I-N ELECTRON JUNCTION DETECTORS
Abstract
The counting performance of p-i-n solid-state detectors for beta and gamma radiation was partially established. Tests with several uncalibrated isotopes and with a calibrated Co(60) source indicated good response to both beta and gamma radiation at several energy levels for a wide range of count rates. The detector obeys the inverse square law and has an efficiency of about 3.6% when counting photons from Co(60). (Author)