METHODS OF MICROANALYSIS V. OXYGEN FLASK COMBUSTION OF CHEMICAL AGENTS
Abstract
The oxygen flask technique of sample combustion was applied to products produced in Chemical Corps research. Evaluation was obtained through selected and modified microvolumetric procedures. Results were obtained for sulfur and halogen elements. The ranges of analytical recovery for each individual element or procedure were 100% ! 0.29% for the average recovery; less than or equal to 0.35 for the average deviation; less than or equal to 0.28 for the standard deviation.A favorable comparison was found by contrasting these quantities with similar ones obtained by other methods. Optimum design parameters for the flask are considered. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Apr 01, 1962
- Accession Number
- AD0278160
Entities
People
- John M. Corliss
Organizations
- Edgewood Chemical Biological Center