THE FERROUS-FERRIC DOSIMETER: A REVIEW,

Abstract

The characteristics, preparation, and use of the ferrous-ferric dosimeter are reviewed in detail. This includes a discussion of the dose range, dose rate dependence, energy dependence, and temperature effects. For photons with energies in excess of 6 kev the irradiation yield may be represented by the expression G = 15.61 - 15.43/E, where E is the mean photon energy in kev and G is the number of ferrous ions oxidized per 100 ev of energy absorbed. Several methods of measuring ferric ion concentration are described, including the conventional spectrophotometric determination of ferric and ferrous ion and the very sensitive radiometric measurement of ferric ion. A concise summary of all the information necessary for routine application of this dosimeter is given at the end of the report. (Author)

Document Details

Document Type
Technical Report
Publication Date
Dec 01, 1963
Accession Number
AD0429156

Entities

People

  • Edward A. Burke

Organizations

  • Air Force Cambridge Research Laboratories

Tags

Communities of Interest

  • Energy and Power Technologies

DTIC Thesaurus Topics

  • Dose Rate
  • Dosimeters
  • Dosimetry
  • Measurement
  • Measuring Instruments
  • Radiation Dosage
  • Radiation Measuring Instruments

Fields of Study

  • Physics

Readers

  • Analytical Chemistry
  • Nuclear and Radiation Engineering.
  • Plasma Physics.