ON IMPEDANCE MEASUREMENTS: THE DETERMINATION OF THE DOUBLE LAYER CAPACITANCE IN THE PRESENCE OF AN ELECTRODE REACTION; AND CAPILLARY RESPONSE AT A DROPPING MERCURY ELECTRODE.
Abstract
A solution film inside the capillary is the principal locus of shielding at a DME. A graphical method for obtaining the true double layer capacitance is presented, and a model is given which represents the electrical characteristics of the solution film. A graphical method is given for the determination of both the double layer capacitance and the faradaic impedance from impedance measurements in the presence of a faradaic process. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Aug 01, 1964
- Accession Number
- AD0604906
Entities
People
- R. De Levie