SPECTRA OF THE INTERNALLY SCATTERED RADIATION FROM LARGE CO 60 SOURCES USED IN TELETHERAPY,
Abstract
Because of internal scattering and degradation of the 1.33 and 1.17 MeV quanta, the beam of radiation from a Co60 teletherapy unit includes a continuous spectrum extending well below 1 MeV. The form of this spectrum below l MeV and its intensity relative to the radiation above 1 MeV have been investigated for sources of various dimensions. The influence of the material around the source, the relative spectral contributions of different parts of the source, and the spectra from bare sources outside the teletherapy head have also been studied. Methods of designing source and head to minimize the radiation below 1 MeV may be inferred from the graphs. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Jan 09, 1964
- Accession Number
- AD0611645
Entities
People
- J. H. Aitken
- W. H. Henry
Organizations
- National Research Council Canada