EVALUATION OF ROBERTSHAW FULTON CONTROLS COMPANY MODEL 2867A1B AND 2867B1 SAMPLE OXYGEN REGULATORS.

Abstract

The regulators submitted by Robertshaw Fulton Controls Company were picked at random from production samples of oxygen breathing regulators model 2867A1B and B1. In accordance with request contained in BUAER 1tr Aer-AE-51/119 of 14 March 1955, these regulators were submitted for evaluation and determination of their compliance with the performance requirements. The results of testing indicate that the regulators do not fully meet the performance requirements with respect to the 'added oxygen' characteristics. This laboratory has recommended that the regulators be considered as unsatisfactory for service use. In addition it is suggested that the manufacturer review his inspection test procedures relative to the inspection testing of 'added oxygen' requirements for the regulators. (Author)

Document Details

Document Type
Technical Report
Publication Date
Mar 21, 1956
Accession Number
AD0627591