CRYOSTAT FOR THE SF-4 SPECTROPHOTOMETER,
Abstract
The article describes in detail the construction of a metallic cryostat for the measurement of optical absorption spectra in the temperature interval 85-300K using the SF-4 spectrophotometer. The cryostat is sufficiently simple and makes it possible to carry out measurements on three samples simultaneously. Not more than 50-60cc of liquid nitrogen is consumed per hour in maintaining the attained temperature. (Author)