METHOD OF DETERMINING THE DROPLET SIZE DISTRIBUTION IN THE ATOMIZATION OF LIQUIDS
Abstract
A laboratory method was developed for determining the droplet size distribution produced by atomizers. The atomizer is placed in a tube in which air is flowing at a velocity of 10 m/sec. The droplets flow downstream around a rod which has one flat surface covered with soot and a thin layer of magnesium. The rod is enclosed in a cylindrical shutter which can be rapidly opened to expose the soot-magnesium surface to the droplets which upon impact leave traces which can be studied microscopically. The instrument was tested with glycerine- water mixtures and an aerosol generator. The droplet distributions obtained were in good agreement with measurements made by two other methods.
Document Details
- Document Type
- Technical Report
- Publication Date
- Sep 05, 1968
- Accession Number
- AD0684661
Entities
People
- N. V. Nikitin
- V. F. Dunskii
Organizations
- National Air and Space Intelligence Center