Pure-Tone Threshold with TDH-49 Earphones
Abstract
Pure-tone thresholds of 24 normal-hearing listeners were determined at 11 audiometric frequencies on two TDH-49 and two TDH-39 earphones. Thresholds were recorded in decibels of sound pressure level, as measured with a National Bureau of Standards type 9-A earphone coupler. Threshold data were used to compute appropriate calibration values for audiometers with TDH-49 earphones, reference Appendix F of ANSI S3.6-1969.
Document Details
- Document Type
- Technical Report
- Publication Date
- Jun 01, 1972
- Accession Number
- AD0744823
Entities
People
- Donald C. Gasaway
- Harrell C. Sutherland Jr.
Organizations
- United States Air Force School of Aerospace Medicine