A Simple Method for Measuring Absolute Diffuse Reflectance with a Laboratory Spectrophotometer.
Abstract
BERG,James D. ;ECOM-5521DA-1-T-061102-B-53-A1-T-061102-B-53-A-19*Diffuse reflection, *Reflectance, Measurement, Spectrophotometers, Barium compounds, SulfatesBarium sulfatesA method is developed for measuring the absolute diffuse reflectance of the reference standard used in making relative diffuse reflectance measurements with an integrating sphere spectrophotometer. No extra experimental set up, such as a second integrating sphere, is required. The method is based on the Kubelka-Munk theory of diffuse reflectance. A graphical procedure is given, and an equation is derived which expresses the absolute reflectance as a function of four readily measurable quantities: two molar concentrations and two relative reflectances. (Modified author abstract)
Document Details
- Document Type
- Technical Report
- Publication Date
- Nov 01, 1973
- Accession Number
- AD0771216
Entities
People
- James D. Lindberg
Organizations
- United States Army Communications-Electronics Command