Evaluation of Differential Amplifier, Microcom Corporation Model MA-20.

Abstract

Laboratory tests were conducted on two light-weight (0.17 ounce) differential amplifiers for use with low-level strain gages. Over the temperature range of -60F to +250F and an elapsed time of 46 days, the combined effects of zero shift and gain variation introduce errors less than plus or minus 3% of full scale. These amplifiers are considered satisfactory for the intended application. (Modified author abstract)

Document Details

Document Type
Technical Report
Publication Date
Dec 12, 1973
Accession Number
AD0774107

Entities

People

  • Sidney Scheindlinger

Organizations

  • Naval Air Warfare Center Warminster

Tags

DTIC Thesaurus Topics

  • Abstracts
  • Amplifiers
  • Corporations
  • Gages
  • Laboratory Procedures
  • Laboratory Tests
  • Strain Gages
  • Test And Evaluation
  • Test Methods

Readers

  • Electronics Engineering
  • Fluid Dynamics.
  • Software Engineering