Evaluation of M15/M18 Enzyme Detector Ticket System with Low Concentrations of GB

Abstract

The objective of this investigation was to evaluate the enzyme ticket system in the M15A2 and M18A2 detector kits, using various types of exposure, for the detection of low concentrations of GB. The tickets met the requirements for the chemical agent detector kit (0.03 plus or minus 0.02 micrograms/l in less than 10 min). They were able to detect 0.02 micrograms/l by static exposure for 8 min.