Polydimethylsiloxane: A Feasibility Study of a Thromboresistant Biomaterial,
Abstract
Silicone rubber as used in blood contacting situations usually contains fillers, catalyst and other additives. In this study, polydimethylsiloxane (PDMS) alone was cross-linked by ionizing radiation in a nitrogen atmosphere to produce a 'pure' silicone rubber. This material exhibited a Lee-White clotting time over 45 minutes and a clotting time of 90 minutes when air-blood contact was minimized. Comparable times for commercial silicones are 25 and 60 minutes. In extracorporeal testing using a 'Mini Lung', an irradiated PDMS coated membrane demonstrated greatly improved thromboresistance over standard silicone rubber membranes. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Jun 01, 1974
- Accession Number
- AD0787464
Entities
People
- E. W. Stool
- P. K. Weathersby
- T. Kolobow
Organizations
- Armed Forces Radiobiology Research Institute