Impulse Assessment of the Honeywell Howard Leight (trademark) MAX (registered trademark) Earplug
Abstract
The impulse peak insertion loss (IPIL) is the standard measure of attenuation provided by hearing protection devices (HPDs) in response to an impulsive noise. This technical memorandum describes the IPIL testing conducted and the calculated mean IPIL values for the Honeywell Howard Leight MAX Earplug (Howard Leight MAX; Model: MAX-1). Testing was in accordance with the ANSI standard S12.42-2010. All samples were tested at the nominal levels of 160 and 170 decibel peak (dBP, re: 20 microPa). A total of five samples were fitted to an acoustic test fixture two times each for a total of 10 trials per test level. No samples of the HPD were rejected. The mean and standard deviation (SD) IPIL values for each nominal level were 47.9 (2.1) dB at 160 dBP and 49.1 (2.1) dB at 170 dBP (see Table 1). These results suggest that when properly fit and functional, the Howard Leight MAX will adequately protect (i.e., reduce exposure to less than 140 dBP) for impulses below 180.0 dBP.
Document Details
- Document Type
- Technical Report
- Publication Date
- Dec 01, 2021
- Accession Number
- AD1153825
Entities
People
- Alexa H. Kolias
- Derek W. Schwaller
- Jeremy S. Federman
- Natalie Silvia
- Stephanie J. Karch
Organizations
- Leidos
- Naval Submarine Medical Research Laboratory