Necessity of Prerinsing Disposable Polypropylene Pipet Tips.
Abstract
Errors as great as 15 percent of rated volume occurred in the first sample dispensed from a piston-actuated microliter pipet using disposable polypropylene tips. Prerinsing of the tip before use in dispensing either aqueous or nonaqueous solutions is recommended to obviate possible errors.
Document Details
- Document Type
- Technical Report
- Publication Date
- Jul 01, 1974
- Accession Number
- ADA003930
Entities
People
- G. H. Zeman
- N. S. Mathewson
Organizations
- Armed Forces Radiobiology Research Institute