This report describes the design and development of a double gap loop B dot sensor in response to a need for a short rise time sensor for pulse radiation field measurements for antenna scale modeling. A sensor with a single-ended or unbalanced output was needed because this was a requirement of the laboratory instrumentation for monitoring fast rise time pulses (on the order of 0.5 nanosecond).