Pulsed Voltammetric Stripping at the Thin Film Mercury Electrode.
Abstract
A computer controlled data acquisition system was used to generate comparative data for thin film anodic stripping voltammetry using staircase, differential pulse and square wave waveforms. Each waveform was tested for its sensitivity and speed of analysis. The square wave waveform was found to be the most sensitive, whether using the square wave or differential pulse measurements. This waveform offers the advantages of fast analysis time and discrimination against charging currents. (Author)