Calibration of Graticules to High Accuracy.

Abstract

A technique for calibrating graticules (or stage micrometers) to high accuracy (+ or - 0.1 micrometers) is described. The lines were sensed photo-electrically and the distance between lines was measured with a laser interferometer. The accuracy is limited mainly by the quality of the graticule lines but for a graticule with long well-defined lines the uncertainty of measurement was 0.04 micrometers. (Author)

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Mar 01, 1977