Improving the Analytical Accuracy of Electron Spin Resonance Spectrometry.

Abstract

Sources of error in Analytical Electron Spin Resonance Spectrometry are discussed. Errors due to gain of the lock in amplifiers and modulation amplitude settings can be reduced if the settings are properly calibrated or if the attenuators are rebuilt. It is also necessary that the incident microwave power level is monitored and that the cavity is reproducibly tuned. A method of correction for incomplete integration of Lorentzian lines is discussed.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Aug 22, 1977