Improving the Analytical Accuracy of Electron Spin Resonance Spectrometry.
Abstract
Sources of error in Analytical Electron Spin Resonance Spectrometry are discussed. Errors due to gain of the lock in amplifiers and modulation amplitude settings can be reduced if the settings are properly calibrated or if the attenuators are rebuilt. It is also necessary that the incident microwave power level is monitored and that the cavity is reproducibly tuned. A method of correction for incomplete integration of Lorentzian lines is discussed.
Document Details
- Document Type
- Technical Report
- Publication Date
- Aug 22, 1977
- Accession Number
- ADA043565
Entities
People
- Ira B. Goldberg