Evaluation of Preproduction Model 3000-Meter Surface-Vessel Bathythermograph OC-6/S,
Abstract
The bathythermograph as tested meets the specification with the following exceptions: the slide-retaining springs are too weak in both the grid holder and the pressure-bourdon slide carriage; the velocity grids are not marked for every 2 meters per second; and the temperature bourdon is not correctly temperature compensated.
Document Details
- Document Type
- Technical Report
- Publication Date
- Oct 24, 1952
- Accession Number
- ADA049840
Entities
People
- R. A. Ross
Organizations
- Navy Electronics Laboratory