Tests on Pneumatically Powered Precision Force Generators.
Abstract
An assessment has been made of the performance of two different types of pneumatic force generator, covering forces up to a 100 kN, with a view to their use in balance calibrations in the RAE 5m wind tunnel. This report describes tests performed on the force generators using a precision pneumatic control system firstly in conjunction with a static weigh beam test rig, followed by a series of tests to check the overall accuracy of performance against the 'dead-weight' Force Standard Machines at NPL Teddington. It has been shown that force generators using a rolling diaphragm seal can produce forces to an accuracy within 0.01% of maximum output assuming a linear calibration. Individual values of force output could be repeated to 0.003% of maximum output when checked against a simple weigh beam under fixed experimental conditions. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Jan 18, 1979
- Accession Number
- ADA085862
Entities
People
- R. D. Law
Organizations
- Royal Aircraft Establishment