US Army Test and Evaluation Command Test Operations Procedure Radiac Ratemeter Calibration Accuracy

Abstract

This is a procedure to standardize methods of determining the calibration accuracy of ratemeters over the range of 80 KEV to 3 MEV. It is to be conducted within a secure enclosure or building where radiation is reduced to a rate less than 2 milliroentgens per hour. Calibration is scored against US army secondary standards. The procedure is used for/with tactical ratemeters.