Calibration of Radiac Calibrator Type AN/UDM-1A.

Abstract

The Radiac Calibrator type AN/UDM-1A has recently been acquired and installed at 9 Canadian Forces Bases across Canada and in Europe. The AN/UDM-1A Calibrator is 120 Ci of 137Cs and provides the required gamma radiation fields for the calibration of CF radiac equipment. This report provides a detailed description of the calibrator and method of the measurement of the gamma fields. (Author)