Determination of Sample Container Effects on Wearmetals Concentration in MIL-L-7808 Synthetic Lubricant Analyzed on an A/E35U-3 Fluid Analysis Spectrometer.
Abstract
This report presents a comparison of three sample containers commonly used with an A/E35U-3 Fluid Analysis Spectrometer. Data from the analysis of wearmetals in MIL-L-7808 synthetic lubricants was analyzed by several standard statistical techniques. These techniques were chosen to detect differences in analytical data and isolate variance that could be attributed to the sample containers. Results of these tests show that the sample containers tested can be used interchangeably and have negligible effects on the analytical method. (Author)
Document Details
- Document Type
- Technical Report
- Publication Date
- Mar 01, 1979
- Accession Number
- ADA107323