Frequency Response Measurements on Lyman-Alpha Humidiometers.

Abstract

The primary instrument used for measuring high-frequency humidity fluctuations is the Lyman-alpha humidiometer; yet, there is virtually no published data on its frequency response. Attempting to fill this void, both amplitude and phase measurements were made on several such instruments. These measurements showed frequency responses (amplitude 3 dB down) ranging from approximately 160 to 800 Hz. Measurements of a specially modified unit showed no fundamental frequency limitation up to at least 10 kHz. (Author)