Measurements of Elongational Viscometry Using a Fiber Spinning Technique. Part 1. Modifications to the Sangamo Schlumberger Elongational Viscometer Model E4,

Abstract

The Sangamo-Schlumberger Elongational Viscometer (Model E4), as delivered by the manufacturer, did not maintain a constant elongational load under constant flow conditions and consequently could not be used to accurately measure elongational viscosity. The extensive modifications made to the instrument at DRES to correct his problem are described in detail as are other improvements made to the instrument.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Feb 01, 1984
Accession Number
ADA140002

Entities

People

  • M. D. Gauthier Mayer
  • S. J. Armour
  • W. J. Fenrick

Organizations

  • Defence Research and Development Canada

Tags

Communities of Interest

  • Advanced Electronics
  • Air Platforms

DTIC Thesaurus Topics

  • Abstracts
  • Body Weight
  • Cameras
  • Construction
  • Fiber Spinning
  • Fibers
  • Flow Rate
  • Materials
  • Measurement
  • Measuring Instruments
  • Photographs
  • Photography
  • Physical Properties
  • Reservoirs
  • Stainless Steel
  • Viscometers
  • Viscosity

Fields of Study

  • Environmental science

Readers

  • Aerospace Test and Evaluation
  • Fluid Dynamics.