The Modification and Application of a Two-Colour Disappearing-Filament Pyrometer to Perform Flame Temperature Measurements.
Abstract
An optical pyrometer of the disappearing-filament type was modified to allow brightness temperature measurements to be made in the green region of the spectrum. The modified instrument was re-calibrated at a wavelength of 547 nm. The manufacturer's calibration of a second pyrometer operating at lambda = 645 nm was verified. Sources of error in the calibrations are discussed. The theory and limitations of two-colour pyrometry as a means of obtaining true source temperatures are presented. Keywords: Micro optical pyrometer; Two color pyrometry; Brightness temperatures; Flame temperatures; Flame emissivity; and Canada.