Determination of Installed Thermocouple Response

Abstract

The validity of the Loop Current Step Response method was established for type J,K,E, and T thermocouples. Prototype equipment configurations and signal processing methods were tested to determine the features needed in a commercial test instrument. This work completes the proof testing of the in-situ method for response time testing of thermocouples as installed in operating processes or test facilities.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Dec 01, 1986