Test Report on the Model 44000 Seismometer System

Abstract

Teledyne Geotech has completed long period subsurface vault tests on four individual 44000 modules; two horizontal and two vertical. Each type of module is tested in tandem by comparing the outputs of aligned sensors. The latest data is outstanding, particularly that of the horizontal. At a period of 20 seconds, both vertical and horizontal channels show incoherent noise power of approximately-180 dB relative to 1 (m/sec-sq)sq/Hz. At a period of 64 seconds, the horizontal data is some 20 dB quieter than any previous borehole data and furthermore shows no increase with period, even at periods as long as 250 seconds. This performance is remarkable for a non-borehole installation. The vertical data shows excellent performance, although some temperature effects seem to be present. A detailed explanation of the experimental history and current tests follows.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Aug 01, 1988
Accession Number
ADA199900

Entities

Organizations

  • Teledyne Technologies

Tags

Communities of Interest

  • Energy and Power Technologies
  • Sensors

DTIC Thesaurus Topics

  • Acquisition
  • Air Force
  • Boreholes
  • Brushless Dc Motors
  • Calibration
  • Data Acquisition
  • Department Of Defense
  • Frequency
  • Friction
  • High Resolution
  • Leveling
  • Materials
  • Noise
  • Seismometers
  • Sliding Friction
  • Spectra
  • Springs

Readers

  • Computational Modeling and Simulation
  • Optical Fiber Sensing and Electromagnetic Propagation.
  • Seismology