Calibration of Nondestructive Birefringence Fuel Thermal Deposit Rating Device

Abstract

This document the results of SwRI comparison of techniques for measuring the thickness of deposits formed on the surface of heated metal tubes during JFTOT testing of fuels. The techniques evaluated in this program included optical birefringence, dielectric strength, and ion milling. A fourth method involving cross sectioning and direct measurement of deposit thickness at high magnification using electron microscopy would serve as referee if wide disagreement among the three primary techniques was encountered. Considering the inherent difficulties in measuring deposits of less than 1 micrometer by any indirect approach, all three methods showed surprisingly good agreement when the data were plotted for comparison. Keywords: Nondestructive testing, Tables data, Graphs.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Aug 24, 1988
Accession Number
ADA206041

Entities

People

  • James G. Barbee
  • L. L. Stavinoha

Organizations

  • Southwest Research Institute

Tags

Communities of Interest

  • Advanced Electronics

DTIC Thesaurus Topics

  • Abstracts
  • Birefringence
  • Calibration
  • Classification
  • Data Sets
  • Dielectric Strength
  • Electron Microscopes
  • Electron Microscopy
  • Measurement
  • Micrometers
  • Microscopes
  • Microscopy
  • Nondestructive Testing
  • Scanning Electron Microscopes
  • Security
  • Tantalum
  • Thickness

Fields of Study

  • Physics

Readers

  • Mechanical Engineering/Mechanics of Materials.
  • Regression Analysis.
  • Thin Film Deposition Science.

Technology Areas

  • Microelectronics
  • Microelectronics - Graphene
  • Microelectronics - Microelectromechanical Systems