Ultrasonic Measurements Research: Progress in 1989
Abstract
The design of a practical and reasonably accurate prototype transfer calibration system for acoustic emission (AE) transducers has evolved in the past year. This year's work involved the use of a 20 cm thick block which replaced the 3.3 cm thick plate that was found to be unsuitable. The prototype transfer calibration system was constructed and demonstrated to be workable. Results from elastic impact experiments were found to be in good agreement with theory. Formulas have been derived for the calculation of Green's functions for multi-layered media, and computer code has been written for calculation of Green's functions for single layered media.
Document Details
- Document Type
- Technical Report
- Publication Date
- Jun 28, 1989
- Accession Number
- ADA234964
Entities
People
- D. G. Eitzen
- F. R. Breckenridge
- N. N. Hsu
- S. E. Fick
- T. M. Proctor
Organizations
- National Institute of Standards and Technology