4 Fluidmaker 2. Test and Evaluation of 6 L/HR Prototype
Abstract
Five hand-operated intravenous (iv) fluidmakers capble of producing 6 L/hr of water for injection have been tested for endotoxin, oxidizables and chloride content and sterility of product water from a challenge water having high levels of all measured parameters. Three of the fluidmakers incorporated a reverse osmosis unit, ion exchange column, water purification filter and sterilizing filter in series; the remaining two lacked the ion exchange column, but were otherwise identical. Fifty 1.0-L samples were produced with each fluidmaker. All samples were sterile and all but one were within the limit for pyrogenicity; 8 percent were out of limit for oxidizables and nearly all were out of limit for chloride ion. Further testing for removal of oxidizables and development of a suitable ion exchange unit are recommended.