Intravenous (4) Fluidmaker 4. A Disposable Device for Preparation of Sterile Water for Injection in a Field Setting
Abstract
A small, disposable system has been developed to prepare USP Sterile Water for Injection (WFI) from a pressurized potable source. The system comprises, in series, a standard laboratory strong acid/strong base ion exchange column, an activated carbon water purification filter, a fine particle filter, a 0.2 um sterilizing filter and a Resuscitation Fluids Production System (REFLUPS) 18 L bag set, which is filled by means of a 19-port distribution head. This system is capable of generating sterile, pyrogen-free water at a fill rate of 0. 5 L/min. Product resistivity did not consistently meet the 1.0 megohm limit corresponding to the USP standard for dissolved salts; a second, smaller, ion exchange cartridge will be required.
Document Details
- Document Type
- Technical Report
- Publication Date
- Jun 01, 1991
- Accession Number
- ADA247385
Entities
People
- James H. Nelson
- Mark O. Schmidt
- W. D. Burrows
Organizations
- United States Army Biological Warfare Laboratories