Test and Evaluation Report of the Physio Control Defibrillator/Monitor Model LIFEPAK (Trademark) 8

Abstract

Physio Control Defibrillator/Monitor Model LIFEPAK 8 was tested for electromagnetic interference/compatibility in the UH-60A helicopter under the U. S. Army Program for Testing Evaluation of Equipment for Aeromedical Operations. The tests were conducted using rent military and industrial standards and procedures for electromagnetic interference/ compatibility and human factors. The LIFEPAK 8 was found to be compatible with U.S. Army medical evacuation UH- 60A Blackhawk. Electromagnetic compatibility, Test and evaluation, Aeromedical equipment.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Dec 01, 1991
Accession Number
ADA248283

Entities

People

  • Bill Olding
  • Jeffrey D. Haun
  • Joseph R. Licina

Organizations

  • United States Army Aeromedical Research Lab

Tags

Communities of Interest

  • Biomedical
  • Energy and Power Technologies
  • Human Systems
  • Space

DTIC Thesaurus Topics

  • Air Force
  • Air Force Facilities
  • Aircraft Equipment
  • Aircrafts
  • Barometric Pressure
  • Biomedical Research
  • Flight Instruments
  • Health Services
  • Human Factors Engineering
  • Measurement
  • Military Research
  • Rotary Wing Aircraft
  • Test And Evaluation
  • Test Equipment
  • Test Facilities
  • Test Methods
  • Warning Systems

Fields of Study

  • Physics

Readers

  • Aviation Safety Risk Assessment.
  • Fault Tolerant Diagnosis of Black and White Balloon Isolation Tests Using ¥.
  • Software Engineering