Test and Evaluation Report of the IVAC (Trade Name) Volumetric Infusion Pump Model 580+EE.

Abstract

The IVAC Volumetric Infusion Pump, Model 580+EE, was tested for electromagnetic interference/compatibility in the UH-60A helicopter under the U. S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interference/compatibility and human factors. The IVAC Model 580+EE did not perform properly in the altitude chamber, vibration test, and excess electromagnetic emissions make it ineligible for an air-worthiness release. Therefore, the unit is not considered compatible with U.S. Army aeromedical aircraft.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
May 01, 1992
Accession Number
ADA251077

Entities

People

  • Bill Olding
  • James E. Bruckart
  • Joseph R. Licina
  • Martin Quattlebaum

Organizations

  • United States Army Aeromedical Research Lab

Tags

Communities of Interest

  • Biomedical
  • Energy and Power Technologies
  • Human Systems
  • Sensors
  • Weapons Technologies

DTIC Thesaurus Topics

  • Air Force
  • Air Force Facilities
  • Aircraft Equipment
  • Aircrafts
  • Altitude
  • Altitude Chambers
  • Artificial Organs
  • Biomedical Research
  • Electromagnetic Interference
  • Health Services
  • Measurement
  • Medical Personnel
  • Military Research
  • Rotary Wing Aircraft
  • Test And Evaluation
  • Test Equipment
  • Test Facilities

Readers

  • Aviation Science / Aeronautics.
  • Civilian Systems Systems Program Capability Development and Upgrade Support Activity Expense and Pay Management.
  • Instructional Design and Training Evaluation.