Test and Evaluation Report of the IVAC (Trade Name) Volumetric Infusion Pump Model 580+EE.
Abstract
The IVAC Volumetric Infusion Pump, Model 580+EE, was tested for electromagnetic interference/compatibility in the UH-60A helicopter under the U. S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interference/compatibility and human factors. The IVAC Model 580+EE did not perform properly in the altitude chamber, vibration test, and excess electromagnetic emissions make it ineligible for an air-worthiness release. Therefore, the unit is not considered compatible with U.S. Army aeromedical aircraft.
Document Details
- Document Type
- Technical Report
- Publication Date
- May 01, 1992
- Accession Number
- ADA251077
Entities
People
- Bill Olding
- James E. Bruckart
- Joseph R. Licina
- Martin Quattlebaum
Organizations
- United States Army Aeromedical Research Lab