Test and Evaluation Report of the IVAC (Trademark) Volumetric Controller Model 280EE

Abstract

The IVAC Volumetric Controller, Model 280EE, was tested for electromagnetic interference/ compatibility in the UH-60A helicopter under the U.S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interference/compatibility and human factors. The IVAC Model 280EE did not perform properly in the altitude chamber and excess electromagnetic emissions make it ineligible for an airworthiness release. Therefore, the unit is not considered compatible with the U.S. Army aeromedical aircraft.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Apr 01, 1992
Accession Number
ADA255919

Entities

People

  • Bill Olding
  • James E. Bruckart
  • Joseph R. Licina
  • Martin Quattlebaum

Organizations

  • United States Army Aeromedical Research Lab

Tags

Communities of Interest

  • Biomedical
  • Energy and Power Technologies
  • Human Systems
  • Sensors
  • Weapons Technologies

DTIC Thesaurus Topics

  • Air Force
  • Air Force Facilities
  • Aircrafts
  • Altitude
  • Altitude Chambers
  • Biomedical Research
  • Electromagnetic Interference
  • Health Services
  • Measurement
  • Medical Personnel
  • Military Research
  • Rotary Wing Aircraft
  • Standards
  • Test And Evaluation
  • Test Equipment
  • Test Facilities
  • Test Methods

Readers

  • Aerospace Test and Evaluation
  • Aviation Science / Aeronautics.
  • Housing Policy Studies in Military Families with Privatization and Telomerase Allowance Units, Multi-Family Housing, and Telomere Lengths.