Ultrahigh Mass Resolution (m/delta m 1/2) > (400,000) Glow Discharge Mass Spectrometry: Direct Analysis of Heavy Isotope Mixtures
Abstract
Fourier transform ion cyclotron resonance (FTICR) mass spectrometry, when coupled to a glow discharge ion source, has demonstrated a mass resolution (m/Delta m sub 1/2) which exceeds that of any currently available instrumentation by a factor of 20. Such high mass resolution is essential for the direct analysis of heavy isotopes without preseparation, where overlapping isobaric peaks, which differ in mass by only a few parts in 500,000, must be resolved.... Glow discharge, Elemental analysis, Heavy isotope analysis, High mass resolution, Fourier transform ion, Cyclotron mass spectrometry.
Document Details
- Document Type
- Technical Report
- Publication Date
- May 05, 1993
- Accession Number
- ADA264377
Entities
People
- C. H. Watson
- C. M. Barshick
- F. H. Laukien
- J. R. Eyler
- J. Wronka
Organizations
- University of Florida