Test and Evaluation Report of the Medical Data Electronics Escort Patient Monitor Model E300A

Abstract

The Medical Data Electronics escort patient monitor, model E300A, was tested for environmental and electromagnetic interference/compatibility in the UH-60A helicopter under the U.S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for environmental tests and electromagnetic interference/compatibility and human factors. The Medical Data Electronics escort patient monitor, model E300A, did not perform properly in the vibration test, high temperature operation test, and low temperature operation test. Excess electromagnetic emissions and electromagnetic susceptibility make it ineligible for an airworthiness release. The unit is not considered compatible with the U.S. Army Medical Evacuation UH-60A Black Hawk

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Mar 01, 1993
Accession Number
ADA267100

Entities

People

  • Brenddan E. Squire
  • James E. Bruckart
  • Leslie W. Johnston
  • Martin D. Quattlebaum

Organizations

  • United States Army Aeromedical Research Lab

Tags

Communities of Interest

  • Advanced Electronics
  • Biomedical
  • Energy and Power Technologies
  • Human Systems

DTIC Thesaurus Topics

  • Air Force
  • Air Force Facilities
  • Biomedical Research
  • Birds
  • Electromagnetic Interference
  • Health Services
  • Liquid Crystal Displays
  • Materials
  • Measurement
  • Medical Personnel
  • Military Research
  • Rotary Wing Aircraft
  • Simulators
  • Test And Evaluation
  • Test Equipment
  • Test Facilities
  • Test Methods

Readers

  • Electrical Engineering
  • Trauma or Military Medicine
  • Women's Health and Cancer Risk Research: African American Women and Pregnancy Outcomes.

Technology Areas

  • Microelectronics
  • Microelectronics - Microelectromechanical Systems