Time-Dependent and Light-Induced Fading in Victoreen Model 2600-80 Aluminum Oxide Thermoluminescence Dosemeters
Abstract
In the absence of light, the thermoluminescence (TL) signal of irradiated aluminum oxide TL dosemeters (TLD) faded by 21% in 3 months. When exposed to low intensity incandescent light. the TL signal decreased by 25% in 6 h. High intensity fluorescent light caused the TL signal to fade by more than 90% in 6 h.
Document Details
- Document Type
- Technical Report
- Publication Date
- Jan 01, 1993
- Accession Number
- ADA268994
Entities
People
- J. H. Musk
Organizations
- Armed Forces Radiobiology Research Institute