Hardware, Software, and Operational Procedures for an Extractive FT-IR Air Monitoring System

Abstract

The report provides a general background for extractive sampling with FT-IR spectroscopy. The report also documents the procedures necessary to validate the performance of an FT-IR spectrometer used in an extractive sampling application. The design, fabrication, and implementation of an extractive FT-IR system are demonstrated. Alignment methods for a 10-m optical gas cell and FT-IR spectrometer is furnished in a step-by-step procedure list. Creation of reference spectra is shown for a target analyte of ethanol at 200 0C. Appendixes contain details on (1) FT-IR spectrometer hardware, (2) sampling protocols for extractive FT-IR spectroscopy, and (3) baseline spectra of ethanol and ethylene simulants.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Apr 01, 1999
Accession Number
ADA364907

Entities

People

  • Charles C. Chaffin
  • Timothy L. Marshall

Tags

Communities of Interest

  • Advanced Electronics
  • Energy and Power Technologies
  • Engineered Resilient Systems
  • Sensors

DTIC Thesaurus Topics

  • Chemistry
  • Circuit Boards
  • Computers
  • Data Analysis
  • Detection
  • Detectors
  • Gas Cells
  • Infrared Detectors
  • Instrumentation
  • Laser Diodes
  • Measurement
  • Printed Circuits
  • Regression Analysis
  • Spectra
  • Spectrometers
  • Spectroscopy
  • Spreadsheet Software

Readers

  • Analytical Chemistry
  • Image Processing and Computer Vision.
  • Software Engineering