Performance Monitoring of Diabetic Patient Systems

Abstract

The minimum variance performance measure is applied to two diabetic patient models under simulated fault scenarios. The Bergman Model and the Automated Insulin Dosage Advisor (AIDA), are controlled in the Internal Model Control (IMC) framework to achieve adequate blood glucose levels. The focus of this paper is to reflect the importance and feasibility of implementing a detection and diagnosis tool such as the minimum variance performance benchmark to an implantable device for diabetics to guarantee adequate control of blood glucose levels.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Oct 25, 2001
Accession Number
ADA411495

Entities

People

  • Camelia L. Owens
  • Francis J. Doyle III

Organizations

  • University of Delaware

Tags

Communities of Interest

  • Materials and Manufacturing Processes

DTIC Thesaurus Topics

  • Absorption
  • Abstracts
  • Artificial Organs
  • Chemical Engineering
  • Control Systems
  • Detectors
  • Dynamics
  • Engineering
  • Equations
  • Meals
  • Military Research
  • Model Predictive Control
  • Monitoring
  • Robots
  • Sensitivity
  • Transfer Functions
  • White Noise

Readers

  • Computational Modeling and Simulation
  • Gulf War Illness and Chronic Multisymptom Illness in Veterans.
  • Logistics and Supply Chain Management.