An Original Method for Non-Invasive Glucose Measurement: Preliminary Results

Abstract

This paper presents first results obtained using a novel approach for non-invasive measurement of glucose content We use a network analyzer, which is used to measure T/R, the modulus of the Transmission to the Reference signal. The under-test solution is placed in the center of a bobbin and plays the role of the self-inductive core. The variation of the "self-inductive medium-core" varies the mutual biductance of the bobbin. We use an electromagnetic signal at a small frequency band about 45 MHz. We have arrived to a measurement about 0.2 g/liter and the results for the temperature influence on the glucose solution are also presented.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Oct 25, 2001
Accession Number
ADA412192

Entities

People

  • A. Rouane
  • M. Gourzi
  • M. Nadi
  • S. M. Alavi

Tags

Communities of Interest

  • Energy and Power Technologies
  • Sensors

DTIC Thesaurus Topics

  • Detection
  • Detectors
  • Electromagnetic Radiation
  • Engineering
  • Flow Rate
  • Frequency
  • Frequency Bands
  • Frequency Response
  • Glucose Monitors
  • Hematologic Tests
  • Instrumentation
  • Ionizing Radiation
  • Measurement
  • Military Research
  • Monitoring
  • Radiation
  • Signal Processing

Readers

  • Computer Engineering
  • Microwave Engineering.
  • Theoretical Analysis.