Abstract

Characteristics and sampling efficiencies of five aerosol samplers designed and built by MesoSyetem Technology, Incorporated (Kennewick, WA) were determined. The samplers tested were three BioCapture(TM) (BT-550) samplers, one MicroVIC(TM) concentrator connected to a filter (MV-F), and one MicroVIC(TM) concentrator connected to an SKC BioSampler (MV-SKC). Monodisperse fluorescent oleic acid particles and monodisperse fluorescent polystyrene latex particles were used in the tests. The analysis was by fluorometry. The results showed that all samplers had a peak sampling efficiency for 2-micrometers particles. The average peak sampling efficiencies of the samplers BT-550, MV-F, and MV-SKC were 38%, 63%, and 27%, respectively.