A PC-Based Time Interval Counter With 200 PS Resolution
Abstract
This paper describes the design and technical features of the precise time-interval counters realized as PC computer boards with ISA, PCI, and PXI interfaces. Thanks to the use of a specialized, interpolating counter chip fabricated in CMOS FPGA technology, the 200 ps resolution (1 LSB - Least Significant Bit) was achieved in single-shot measurements of time intervals within the range from 0 to 43 s. The standard measurement uncertainty is below 1 LSB as a result of automatic correction of nonlinearity of two interpolators contained in the FPGA counter chip. The counter board can also be used for frequency measurement up to 1.1 GHz. The dedicated software operating in a Windows environment provides comprehensive control, diagnostics, and statistical processing of the measured data.
Document Details
- Document Type
- Technical Report
- Publication Date
- Sep 01, 2004
- Accession Number
- ADA427911
Entities
People
- Jozef Kalisz
- Ryszard Szplet
Organizations
- Military University of Technology