Towards an Assurance Case Practice for Medical Devices

Abstract

Software technology enables an increasing percentage of medical device functionality, leading to much more complex systems and presenting a challenge to regulators charged with evaluating device safety and effectiveness. An approach to evaluating claims of safety increasingly used in Europe and elsewhere is the safety assurance case. Much like a legal case, the assurance case lays out an argument and supporting evidence to show that safety claims are valid. This technical note explores the use of assurance cases for justifying claims of medical device safety. It illustrates the use of the assurance case on a particular type of medical device-the infusion pump. This example serves as a basis for discussing issues surrounding the introduction of assurance cases into the medical device community, which includes both manufacturers and the U.S. Food and Drug Administration.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Oct 01, 2009
Accession Number
ADA512366

Entities

People

  • Charles Weinstock
  • John B. Goodenough

Organizations

  • Carnegie Mellon University

Tags

Communities of Interest

  • Biomedical

DTIC Thesaurus Topics

  • Artificial Organs
  • Cardiovascular System
  • Communities
  • Computer Programming
  • Computer Science
  • Department Of Defense
  • Engineering
  • Health Care
  • Health Services
  • Infusions
  • Medical Personnel
  • Software Design
  • Software Development
  • Standards
  • United States
  • Urban Areas
  • User Interface

Readers

  • Aviation Safety Risk Assessment.
  • Strategic Security Studies