Vapor Pressure Determination of VM Using the Denunder-Liquid Chromatography-Mass Spectrometry Technique

Abstract

The vapor pressure of O-ethyl S-2-(diethylamino) ethyl methylphosphonothiolate (VM) has been determined using a novel, denuder-based liquid chromatography mass spectrometry (LC MS) approach designed specifically for low volatility and thermally unstable materials. This system incorporates a novel, packed-bed, vapor-generation system specifically designed to deliver low-volatility and thermally unstable compounds at low or high flow rates using milligram quantities of the material. The vapor pressure analysis system takes advantage of denuder technology to efficiently transfer the chemical sampled from the airstream into a liquid vehicle for transport to a detector such as LC MS. Capturing the chemical of interest into a liquid vehicle eliminates the need for heated transfer lines by virtually eliminating line losses. The results of the vapor pressure study of VM are presented.

Open PDF

Document Details

Document Type
Technical Report
Publication Date
Jan 01, 2015
Accession Number
ADA613614

Entities

People

  • Ai Sohrabi
  • Amanda L. Jenkins
  • Eric J. Bruni
  • Leonard C. Buettner
  • Michael W. Ellzy

Organizations

  • Edgewood Chemical Biological Center

Tags

Communities of Interest

  • Counter WMD

DTIC Thesaurus Topics

  • Chemical Analysis
  • Chromatography
  • Department Of Homeland Security
  • Detectors
  • Flow Rate
  • Fluoropolymers
  • Gas Chromatography
  • Liquid Chromatography
  • Mass Spectrometry
  • Materials
  • Measurement
  • Physical Properties
  • Pressure Measurement
  • Spectrometry
  • Standards
  • Vapor Pressure
  • Volatility

Fields of Study

  • Chemistry

Readers

  • Analytical Chemistry
  • Combustion and Flow Dynamics.