Antimicrobial Efficacy of External Fixator Pins Coated with a Lipid Stabilized Hydroxyapatite/Chlorhexidine Complex to Prevent Pin Tract Infection in a Goat Model
Abstract
Background: Pin tract infection is a common complication of external fixation. An antiinfective external fixator pin might help to reduce the incidence of pin tract infection and improve pin fixation. Methods: Stainless steel and titanium external fixator pins, with and without a lipid stabilized hydroxyapatite/chlorhexidine coating, were evaluated in a goat model. Two pins contaminated with an identifiable Staphylococcus aureus strain were inserted into each tibia of 12 goats. The pin sites were examined daily. On day 14, the animals were killed, and the pin tips cultured. Insertion and extraction torques were measured. Results: Infection developed in 100% of uncoated pins, whereas coated pins demonstrated 4.2% infected, 12.5% colonized, and the remainder, 83.3%, had no growth ( p less than 0.01). Pin coating decreased the percent loss of fixation torque over uncoated pins ( p = 0.04). Conclusion: These results demon- strate that the lipid stabilized hydroxyapatite/chlorhexidine coating was successful in decreasing infection and improving fixation of external fixator pins.