Z-scan Measurement of Upconversion in Er:YAG
Abstract
We have measured the cooperative upconversion parameter in Er:YAG using the z-scan technique as a function of Er concentration (0.5% to 5%) with the sample at room temperature (298 K) and 80 K. The value we obtain at room temperature is 2-3 times larger than previously reported. The value we obtain at 80 K is approximately eight times larger than the value at room temperature. One advantage of the z-scan technique is that the measurement can be carried out with the same quasi-cw diode laser that will pump the Er:YAG laser being developed.
Document Details
- Document Type
- Technical Report
- Publication Date
- Dec 01, 2010
- Accession Number
- ADA639831